VECTOR™ S310 INDICATOR SPECIFICATIONS:
Model
S310 S310D
4 Digit LCD with Selectable
Analog Meter Movement
Full Scale Ranges with Astral 25mm Calorimeter
Full Scale Ranges with Astral 50mm Calorimeter
Full Scale Ranges with Vector Pyroelectric Detector
Full Scale Ranges with Vector HR Pyroelectric Detector
Full Scale Ranges with Ultra Calorimeter
Maximum Repetition Rate with Calorimeter in Joules Mode
Maximum Repetition Rate with Calorimeter in Watts Mode
Maximum Repetition Rate for Collecting Data in
Statistics Mode with a Pyroelectric Detector
Response Time with Calorimeter in Joules Mode
Response Time with Calorimeter in Watts Mode
Response Time with Ultra Calorimeter
Operation Temperature
Power Requirement
Dimensions H x W x D - inches/cm
Weight - pounds/kgs
* 3.000
µ
and 30.00
µ
ranges not available for long pulse setting with PHF02, PHF05 or PHF09 HR Pyroelectric detectors
Display
4 Digit LCD
10.00 m, 100.0 m, 1.000, 10.00, AUTO (Watts only)
Calorimeter Dependent - 1 pulse every 60 to 90 seconds
Unlimited
750 pps
300.0 m, 3.000, 30.00, AUTO (Watts only)
3.000 m, 30.00 m, 300.0 m, 3.000, 30.00, AUTO
3.000 µ*, 30.00 µ*, 300.0 µ, 3.000 m, 30.00 m, AUTO
150.0 Watts
120 Volts, 60 Hz
±
10 % or 220 Volts, 50 Hz
±
10 %
4.68 x 8.83 x 7.83/11.89 x 22.43 x 19.89
5/2.2
Calorimeter Dependent - 1 to 3 seconds
Calorimeter Dependent - 3 to 10 seconds
40 seconds
5
º
C to 40
º
C
ABSORPTION OF HD ABSORBING MATERIAL:
Warning: You must exercise caution when using HD detectors. They exhibit spectral reflection of
between 7% and 18%, of the input power, back out of the aperture. Please refer to
Figure 1 to determine the reflectance for the wavelength you are measuring. These
detectors should be treated as a partial mirror or any other type of reflective optic and
the appropriate caution level observed, especially at the CO
2
wavelength.
ABSORPTION OF HD ABSORBER Vs. W AVELENGTH
50
55
60
65
70
75
80
85
90
95
100
0110100
W AVELEN G TH [M ICRO M ETERS]
ABSORPTION [%]
CAUTION: SPECULAR REFLECTION
Figure 1
Note: HD detectors might show “beam” marks on the absorbing surface. These marks are
characteristic of the material and do not affect the performance of the detector. Do not rub off or
remove the marks. Polishing or cleaning the absorbing surface might change the performance of
the pyroelectric detector.